Each Unit Size (W*D*H)(mm)W*D*H:443*615*125(mm)
-
AC InputAC 220V +10%/ -20%, 50Hz;
-
Input Active Power500W
-
ResolutionAD:16bit;DA:16bit
-
Input Impedance≥ 500MΩ 开机状态(关机时漏电流为100uA)
-
Voltage
Voltage Range per Channel
充电:0.7 V~5V;放电:0.7 V~5V(0.5m夹具线长);
Minimum Discharge Voltage
0.7V(0.5m夹具线长);
Accuracy
± 0.1% of FS;
Stability
± 0.005% of FS;
-
Current
Output Range per Channel
Range 1: 0.1μA~150μA;
Range 2: 150μA~5mA;
Range 3: 5mA~150mA;
Range 4: 150μA~5A;
Accuracy
± 0.02% of FS;
Constant Voltage Cut-off Current
Range 1: ±30nA;
Range 2: ±1μA;
Range 3: ±30μA;
Range 4: ±1mA;
Stability
± 0.005% of FS;
-
Power
Single Channel Maximum Output Power: 25W;
Stability: ± 0.01% of FS;
-
Time
Current Response Time (10%FS~90%FS)
≤100μs
Step Time Range
≥10ms
-
Data Acquisition
Data Acquisition Conditions
Time△t: ≥10ms;
Voltage△U: ≥1mV;
Current△I: ≥100nA;
Data Acquisition Frequency
1000Hz (Continuous Charge and Discharge Mode);
-
Charge
Charging Mode: Constant Current Charging, Constant Voltage Charging, Constant Current/ Constant Voltage Charging, Constant Power Charging;
Cut-off Conditions: Voltage, Current, Relative Time, Capacity, Energy, Power;
-
Discharge
Discharge Mode: Constant Current Discharge, Constant Power Discharge and Constant Resistance Discharge;
Cut-off Conditions: Voltage, Current, Relative Time, Capacity;
-
Cycle Life
Cycle Test Range: 1~65535 times;
Number of Single Cycle Steps: 255;
Loop Nesting: It has the Function of Nested Loop and Supports Up to 4 Levels of Nesting;
-
DCIR TestSupport DCIR Step.
-
Protect
Power down Data Protection;
The safety protection conditions can be set, and the setting specs include: Voltage Upper Limit, Voltage Lower Limit, Current Upper Limit, Current Lower Limit, Capacity Upper Limit Protection, Voltage Fluctuation Protection, Current Fluctuation Protection;
Function of Offline Test.
-
Channel Characteristics4 Range, Wide Dynamic Range, High Speed Sampling, High Precision
-
Channel Control ModeIndependent Control
-
Voltage and Current SamplingFour Wire Connection
-
DatabaseAdopting MySQL database to centralized manage testing data.
-
Communication Mode of Upper ComputerBased on TCP/ IP Protocol (100M Ethernet)
-
Server Disk Configuration500GB
-
Server Operating SystemWindows 7 64Bit
-
Leakage Current0.005mA
-
Data Output Mode
EXCEL 2003, 2010;
TXT.
-
Communication InterfaceNetwork Interface
-
Total Number of Channels8
-
Working Temperature Range25±5℃(精度保证),25±20℃(极限使用温度);
-
Storage Temperature Range0~60℃
-
Relative Humidity Range of Working Environment≤70% RH(There is no condensation of moisture.)
-
Relative Humidity Range of Storage Environment≤80% RH(There is no condensation of moisture.)