Each Unit Size (W*D*H)(mm)
-
AC InputAC 220V ±10% / 50Hz
-
Input Active Power2127W
-
ResolutionAD:16bit;DA:16bit
-
Input Impedance≥1kΩ
-
Voltage
Constant Voltage Range Control: 0.025V~5V;
Minimum Discharge Voltage: 2.5V;
Accuracy: ± 0.1% of FS;
Stability: ± 0.1% of FS;
-
Current
Output Range per Channel: 0.15A~30A;
Accuracy: ± 0.1% of FS;
Constant Voltage Cut-off Current: 0.06A;
Stability: ± 0.1% of FS;
-
Power
Single Channel Maximum Output Power: 150W;
Stability: ± 0.2% of FS;
-
Time
Current Response Time (10%FS~90%FS)
Maximum current rise time: 20ms;
Step Time Range
① ≤(365*24) h/step; ②Time Format Support 00:00:00:00(h, min, s, ms)
-
Data Acquisition
Data Acquisition Conditions
Minimum Time Interval: 100ms;
Minimum Voltage Interval: 10mV;
Minimum Current Interval: 60mA;
Data Acquisition Frequency
10Hz;
-
Charge
Charging Mode: Constant Current Charging, Constant Voltage Charging, Constant Current/ Constant Voltage Charging, Constant Power Charging;
Cut-off Conditions: Voltage, Current, Relative Time, Capacity, -△V;
-
Discharge
Discharge Mode: Constant Current Discharge, Constant Voltage Discharge, Constant Power Discharge and Constant Resistance Discharge;
Cut-off Conditions: Voltage, Current, Relative Time, Capacity;
-
Pulse Mode
Charging: Constant Current Mode and Constant Power Mode;
Discharge: Constant Current Mode and Constant Power Mode;
Minimum Pulse Width: 500ms;
Number of Pulses: A Single Pulse Step Supports 32 Different Pulses;
Continuous Switching of Charge and Discharge: One Pulse Step can Realize Continuous Switching from Charge to Discharge;
Cut off Conditions: Voltage, Relative Time;
-
Cycle Life
Cycle Test Range: 1~65535 times;
Number of Single Cycle Steps: 254;
Loop Nesting: It has the Function of Nested Loop and Supports Up to 3 Levels of Nesting;
-
DCIR TestSupport DCIR calculating with user defined point.
-
Protect
Power down data protection;
Function of prevention of feedback;
The safety protection conditions can be set, and the setting specs include: voltage upper limit, voltage lower limit, current upper limit, current lower limit and delay time;
Function of offline test.
-
IP Protection DegreeProtection Degree: IP20
-
Channel CharacteristicsConstant current and constant voltage adopt double closed loop structure.
-
Channel Control ModeIndependent Control
-
Voltage and Current SamplingFour Wire Connection
-
Noise<85dB
-
DatabaseAdopting MySQL database to centralized manage testing data.
-
Server Disk Configuration500GB
-
Communication Mode of Upper ComputerBased on TCP/ IP Protocol
-
Leakage Current0.005mA
-
Server Operating SystemWindows 7
-
Data Output Mode
EXCEL 2003, 2010;
TXT.
-
Communication InterfaceNetwork Interface
-
Total Number of Channels8
-
Working Temperature Range0℃~40℃(Within the range of 25±10℃, accuracy drift is 0.005% of FS /℃)
-
Storage Temperature Range-10℃~50℃
-
Relative Humidity Range of Working Environment≤70% RH(There is no condensation of moisture.)
-
Relative Humidity Range of Storage Environment≤80% RH(There is no condensation of moisture.)